| Parameter |
Range |
Best
Uncertainty |
Note* |
| * S - Source only, M - Measurement
only, S/M - Source and Measurement |
| DC
VOLTAGE |
100 uV to 30 kV |
+/- 0.0002 % |
S/M |
| Specific
values |
1 mV |
+/- 0.02 % |
S/M |
| 10 V |
+/- 0.0002 % |
S/M |
| 50 kV |
+/- 0.02 % |
S/M |
| |
| AC
VOLTAGE |
100 mV to 1 kV @ DC to 1MHz ** |
+/- 0.002 % |
S/M |
| Specific
values |
1 mV / 1 kHz |
+/- 0.1 % |
S/M |
| 1 V / 1 kHz |
+/- 0.002 % |
S/M |
| 5 kV / 60 Hz |
+/- 0.1 % |
S/M |
| 30 kV / 60 Hz |
+/- 2 % |
M |
| ** Not all voltage/frequency
combinations available |
| DC
CURRENT |
1 pA to 100 A |
+/- 0.001 % |
S/M |
| Specific
values |
1 pA |
+/- 0.5 % |
S/M |
| 1 mA |
+/- 0.001 % |
S/M |
| 100 A single turn |
+/- 0.1 % |
S/M |
| |
| AC
CURRENT |
10 uV to 100 A @ DC to 10 kHz ** |
+/- 0.02 % |
S/M |
| Specific
values |
10 uA / 1kHz |
+/- 0.04 % |
S/M |
| 1 mA / 1 kHz |
+/- 0.02 % |
S/M |
| 100 A / 60 Hzsingle turn |
+/- 0.1 % |
S/M |
| ** Not all voltage/frequency
combinations available |
| RESISTANCE |
1 mOhm to 8 TOhm |
+/- 0.0002 % |
S/M |
| Specific
values |
1 mOhm |
+/- 0.05 % |
S/M |
| 10 kOhm |
+/- 0.0002 % |
S/M |
| 8 TOhm |
+/- 0.5 % |
S/M |
| |
| CAPACITANCE |
1 pF to 1 F @ 20 Hz to 100 kHz |
+/- 0.0002 % |
S/M |
| Specific
values |
1 pF / 1 kHz |
+/- 0.001 % |
S/M |
| 100/10 pF |
+/- 0.0002 % |
S/M |
| 10 mF / 100 Hz |
+/- 0.005 % |
S/M |
| 1 F / 100 Hz *** |
+/- 0.25 % |
S/M |
| *** 4 Terminal Ratio Type |
| INDUCTANCE |
100 uH to 10 H |
+/- 0.02 % |
S/M |
| |
| FREQUENCY |
DC to 26 GHz |
+/- 5E-11 |
S/M |
| |
| RF VOLTAGE 50 Ohm |
0.5 to 3 V @ up to 100 MHz |
+/- 0.02 % |
S/M |
| Specific
values |
1 V / 1 MHz |
+/- 0.02 % |
S/M |
| 1 V / 100 MHz |
+/- 0.7 % |
S/M |
| |
| RF POWER |
1 mW / 50 MHz |
+/- 0.7 % |
S/M |
| 10 uW to 100 mW / 100 KHz to 26 GHz |
+/- 0.1 % |
S/M |
| 80 W / 20 MHz to 1 GHz |
+/- 3 % |
S/M |
| |
| RF
ATTENUATION |
0 to 120 dB @ 2.5 to 1300 MHz |
+/- 0.05 dB |
S/M |
| |
| TEMPERATURE |
0 to 600 C |
+/- 0.01 C |
S/M |
| |
| SOUND
(spl) |
74 to 124 dB |
+/- 0.2 dB |
S/M |
| |
| OPTICAL POWER |
-70 dBm to 20 dBm @ 1300 / 1500 nm |
+/- 5 % |
M |
| -60 dBm to 0 dBm @ 1300 / 1500 nm |
+/- 5 % |
S |